Evaluation of dosimetric characteristics of a grid block fabricated for Mega-voltage grid therapy purposes
چکیده انگلیسی مقاله
Background: In conventional radiation therapy, regarding normal tissue tolerance, the treatment of bulk tumors is one of the remaining challenges. Grid Radiation Therapy (GRT) is a technique to deliver high doses, approximately 15 – 20 Gy per fraction, to several small volumes located in a large radiation field. This can be performed using a grid block. The current work has concentrated on the dosimetric characteristics of a designed mega-voltage grid, used for a unique treatment modality. Materials and Methods: All measurements performed using a Neptune linear accelerator (9 MV photon beam). A square 16 × 16 array grid block was designed and fabricated. Several dosimetric characteristics including: depth dose, Valley To Peak (VTP) ratio, and grid out-put factor were evaluated using a calibrated diode dosimeter for a range of radiation fields. Results: The percent depth dose curves, measured in the presence of grid block, lie within those measured for the corresponding open field and a narrow beam. At the Dmax, the VTP ratio was found to be within 17% - 28%, while these ranges between 23% - 35% at a depth of 10 cm. The grid out -put factor found to be 0.78 and it slightly decreases with increasing of radiation field size. Conclusion: The VTP ratio found to be dependent strongly on the grid design and manufacturing properties. However, other parameters such as radiation field size and the depth of measurement should also be addressed as important factors. The measured dosimetric characteristics of grid block indicate that the mega-voltage grid therapy can be applied as a possible clinical modality for palliative cases. Iran. J. Radiat. Res., 2011 9(1): 49-56
کلیدواژههای انگلیسی مقاله
Grid therapy, palliation, mega-voltage, radiation dosimetry.
نویسندگان مقاله
k قاضی خانلو ثانی | k ghazi khanlou sani
department of radiology, hamadan university of medical sciences, hamadan, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی همدان (Hamadan university of medical sciences)
m محمدی | m. mohammadi
department of medical physics, hamadan university of medical sciences, hamadan, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی همدان (Hamadan university of medical sciences)
a اسکندرلو | a. eskandarlo
department of radiology, hamadan university of medical sciences, hamadan, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی همدان (Hamadan university of medical sciences)
h غلام حسینیان | h gholam hosseinian
department of radiotherapy, mash-had university of medical sciences, mash-had, iran
نشانی اینترنتی
http://www.ijrr.com/browse.php?a_code=A-10-1-384&slc_lang=en&sid=en
فایل مقاله
فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
Radiation Biology
نوع مقاله منتشر شده
تحقیق بدیع
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات