این سایت در حال حاضر پشتیبانی نمی شود و امکان دارد داده های نشریات بروز نباشند
International Journal of Radiation Research، جلد ۲۰، شماره ۱، صفحات ۱۸۵-۱۸۹

عنوان فارسی
چکیده فارسی مقاله
کلیدواژه‌های فارسی مقاله

عنوان انگلیسی Radiation dose of head and abdomen-pelvis computed tomography examinations using size-specific dose estimate
چکیده انگلیسی مقاله Background: Concern about radiation risk of computed tomography (CT) scan as a diagnostic modality has increased in recent years. Diagnostic reference levels (DRLs) is one of the tools to optimizing radiation dose of patients. CTDIv (Volume Computed tomography dose index) and DLP (Dose Length product) are used for assessment of DRLs. The CTDIv under/overestimate the patient dose. AAPM has introduced SSDE (Size-specific dose estimates) for estimation of patient. In this study, the DRLs of head and abdomen-pelvis CT examinations of adults is determined using CTDIv, DLP and SSDE. Materials and Methods: 680 CT examinations of head and abdomen-pelvis were collected from PACS (Picture archiving and communication system) in Imam Khomeini and Mostafa Khomeini hospitals. The Deff, CF and SSDE calculated using AAPM TG-204 and TG-220. Statistics analysis calculated using SPSS version 18. Results: For abdomen-pelvis third quartile of CTDIV, SSDE and DLP was 9.96, 13.58 and 527 and values of 27.62, 26.79 and 402.90 are determined for head, respectively that are lower than national DRLs. Also, calculated conversion factor (CF) for head and abdomen-pelvis was 0.97 ± 0.75and 1.45 ± 0.17, respectively. Conclusion: DRLs were lower than other studies in this study. Using the AEC (Auto Exposure Control) and different kVp in this hospitals can help optimization of patient dose. The SSDE must be calculable by radiographers to more accurate estimation of patient dose using CFs.  
کلیدواژه‌های انگلیسی مقاله Head imaging, neuroradiology, abdominal imaging, SSDE, TG 204, TG 220.

نویسندگان مقاله | M. Ahmadifard
Radiobiology and Radiation Protection Dept., School of Allied Medical Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran


| M. Bakhshandeh
Department of Radiology Technology, Allied Medical Faculty, Shahid Beheshti University of Medical Sciences, Tehran, Iran


| A. Mohammadbeigi
Students Research Committee, Kermanshah University of Medical Sciences, Kermanshah, Iran


| K. Khoshgard
Department of Medical Physics, School of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran



نشانی اینترنتی http://ijrr.com/browse.php?a_code=A-10-2188-80&slc_lang=en&sid=1
فایل مقاله فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده en
موضوعات مقاله منتشر شده Radiobiology
نوع مقاله منتشر شده تحقیق بدیع
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات