عنوان انگلیسی The Dosimetric Comparison between Measured Tissue Maximum Ratio Directly by Water Phantom and That Calculated from Percentage Depth Dose Measurments in Small Field and Determind the Differnces between Two Methods
چکیده انگلیسی مقاله Introduction: The aim is to observe that the measurement of tissue maximum ratio (TMR) by water phantom directly differs from the calculation of TMR from percentage depth dose (PDD) or not.Material and Methods: The linear accelerator Siemens (6 MV & 10MV) with 82 leaves (MLC)-based stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). The water phantom (PTW) was used for measuring the dosimetric parameter of TMR &%DD. This data was calculated using 0.125 cc simeflex ionization chamber. The small fields have different sizes ranging from 12.50 to 40.00 mm in diameter.Results: There were small observations of mean error ≤ 1.50 % for all collecting data related to depths and field sizes. The Statistical Package of Social Science (SPSS) (version 26) to generate results. Wilcoxon Signed Ranks Test was used to compare two groups; (p ≤ 0.05) was considered significant. There were no significant differences between TMR data. The measured TMR data calculated from %DD had a strong positive correlation for cone sizes from 1.0 cm x1.0 cm to 10.0 cm x10.0cm.Conclusion: It has been shown that the calculation of TMR data from PDD agrees with the measuring values directly, and it is accepted for use in treatment.
کلیدواژه‌های انگلیسی مقاله Measured TMR, Calculated TMR, Small field dosimetry

نویسندگان مقاله | Ehab Attalla
Physics Unit-Radiation Oncology, Children’s Cancer Hospital, Cairo, Egypt


| Rasha Elawady
Physics Unit-Radiation Oncology, Children’s Cancer Hospital, Cairo, Egypt


| Shimaa shoier
Department of Radiotherapy & Nuclear Medicine, National Cancer Institute, Cairo University, Cairo, Egypt



نشانی اینترنتی https://ijmp.mums.ac.ir/article_16977.html
فایل مقاله فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده Original Paper
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات