این سایت در حال حاضر پشتیبانی نمی شود و امکان دارد داده های نشریات بروز نباشند
صفحه اصلی
درباره پایگاه
فهرست سامانه ها
الزامات سامانه ها
فهرست سازمانی
تماس با ما
JCR 2016
جستجوی مقالات
شنبه 6 دی 1404
Iranian Journal of Medical Physics
، جلد ۱۸، شماره ۵، صفحات ۳۰۰-۳۰۵
عنوان فارسی
چکیده فارسی مقاله
کلیدواژههای فارسی مقاله
عنوان انگلیسی
Energy Window Optimization Using Triple Energy Window Scatter Correction Method for In-111 SPECT Imaging: A SIMIND Monte Carlo Study
چکیده انگلیسی مقاله
Introduction: Detecting scattered photons in the photo peak window degrades the image contrast and quantitative accuracy of single-photon emission computed tomography (SPECT) imaging. This study aimed to determine optimal main- and sub-energy windows for Triple Energy Window (TEW) in In-111. Material and Methods: We used the simulating medical imaging nuclear detectors (SIMIND) program to simulate the Siemens SYMBIA gamma camera equipped with a medium energy (ME) collimator. We also used the SIMIND Monte Carlo program to generate theIn-111SPECT projection data of the Jaszczak phantom. The phantom consisting of six spheres with different diameters (9.5, 12.7, 19.1, 15.9, 25.4, and 31.8 mm) was used to evaluate the image contrast. Geometric, scatter, and penetration fraction, point spread functions, and contrast curves were drawn and compared. Results: The results showed that the 171keVphotopeak compared to the 245keVphotopeak yielded the best results with an ME collimator when the TEW scatter correction method was applied. The reason can be the large amount of scatter and penetration from the photo peak and the collimator for the 245keVphotopeak window. Conclusion: With the TEW scatter correction method, it is better to use a 171keVphotopeak window because of its better spatial resolution and image contrast.
کلیدواژههای انگلیسی مقاله
SPECT Contrast In, 111 SIMIND Program
نویسندگان مقاله
| Hicham Asmi
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco
| Farida Bentayeb
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco
| Youssef Bouzekraoui
Higher Institute of Health Sciences Settat, University HASSAN, Rabat, Morocco
| Faustino Bonutti
Academic Hospital of Udine, Medical Physics Department, Italy
نشانی اینترنتی
https://ijmp.mums.ac.ir/article_16351.html
فایل مقاله
فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
Original Paper
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات