Analysis of Commissioning Parameters and its Validation of O-Ring Gantry Based Medical Linear Accelerator HalcyonTM for Improved Radiotherapy Technique
چکیده انگلیسی مقاله
Introduction: Popularly, teletherapy (telecobalt/LA) equipment is based on a C-arm gantry system. Recently, a fast O-ring gantry system introduced a medical linear accelerator (LA) to smoothen the workflow of treatment of cancer patients because of the increasing trend of the number of cancer cases over the past few years. This study aimed to analyze the commissioning parameters and validation of the O-ring gantry-based LA for improved radiotherapy techniques. Material and Methods: Three-dimensional (3D) radiation field analyzer (RFA) used to commission HalcyonTM LA. It is used for measuring percent depth dose (PDD), profiles, and output factors. Results: TPS data was validated by comparing it with our measured data. Plans per the TG-119 protocol showed good agreement between treatment planning systems (TPS) calculated and measured doses. For patient-specific, QA plans showed good agreement with gamma evaluation criteria. Conclusion: Commissioning and validation of O-ring gantry system HalcyonTM LA was performed successfully.
کلیدواژههای انگلیسی مقاله
Linear accelerator O, ring gantry Commissioning Validation Radiotherapy technique
نویسندگان مقاله
| ARVIND KUMAR
Graphic Era University, Dehradun, Uttarakhand, India 248002 2. Department of Radiation Oncology, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India 249203
| KIRAN SHARMA
Graphic Era University, Dehradun, Uttarakhand, India 248002
| CHANDI BHATT
Sarvodaya Hospital & Medical Research Centre, Faridabad (Delhi/NCR), India 121006
| Suresh Yadav
Department of Radiation Oncology, Gandhi Medical College , Bhopal-462001 (M.P.), India
| SWATI VERMA
Department of Radiation Oncology, All India Institute of Medical Sciences (AIIMS), Rishikesh, Uttarakhand, India 249203
| SOUGATA BASU
Department of Radiation Oncology, All India Institute of Medical Sciences (AIIMS), Rishikesh, Uttarakhand, India 249203
| POOJA DHAUNDIYAL
Department of Radiation Oncology, All India Institute of Medical Sciences (AIIMS), Rishikesh, Uttarakhand, India 249203
| DEEPAK KAUSHIK
Department of Radiation Oncology, All India Institute of Medical Sciences (AIIMS), Rishikesh, Uttarakhand, India 249203
| DINESH SAROJ
Chirayu Cancer Hospital Radiotherapy Depatment Chirayu Medical College Hospital Bhopal Near Bairagarh
| Manoj Gupta
Department of Radiation Oncology, All India Institute of Medical Sciences (AIIMS), Rishikesh, Uttarakhand, India 249203
نشانی اینترنتی
https://ijmp.mums.ac.ir/article_21991.html
فایل مقاله
فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
Original Paper
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات