Iranian Journal of Medical Physics، جلد ۲، شماره ۳، صفحات ۶۱-۶۵

عنوان فارسی
چکیده فارسی مقاله
کلیدواژه‌های فارسی مقاله

عنوان انگلیسی Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences
چکیده انگلیسی مقاله Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of non-homogeneity in RF receive and transmit. This can be the most effective source of error in quantitative studies in MRI imaging. Part of this non-homogeneity demonstrates the characteristics of RF coil and part of it is due to the interaction of RF field with the material being imaged. In this study, RF field non-homogeneity of surface and volume coils is measured using an oil phantom. The method employed in this work is based on a routine Spin Echo based sequence as proposed by this group previously. Materials and Methods: For the determination of RF non-uniformity, a method based on Spin Echo sequence (θ-180) was used as reported previously by the same author. In this method, several images were obtained from one slice using different flip angles while keeping all other imaging parameters constant. Then, signal intensity at a ROI from all of these images were measured and fitted to the MRI defined mathematical model. Since this mathematical model describes the relation between signal intensity and flip angle in a (θ-180) Spin Echo sequence, it is possible to obtain the variation in receive and transmit sensitivity in terms of the variation of signal intensity from the actual expected values. Since surface coils are functioning as only receiver (RF transmission is done by Body coil), first the results of receive coil homogeneity is measured, then characteristic of transmit coil (for the body coil) is evaluated Results: The coefficient of variation (C.V.) found for T(r) value obtained from images using head coils was in the order of 0.6%. Since the head coil is functioning as both transmitter and receiver, any non-uniformity in either transmit or receive stage can lead to non-homogeneity in RF field. A part from the surface coils, the amount of non-homogeneity due to receive coil was less than that of the transmit coil. In the case of the surface coils the variation in receive homogeneity measurement was high due to inherent sensitivity of the coil in different locations, which was as expected. Discussion and Conclusion: Since the RF interaction of RF field is very low in oil phantom, the variation shown in the T(r) is due to B1 non-homogeneity of coil system. Since the head coil is functioning as both transmitter and receiver, any non-uniformity in either transmit or receive stage could lead to non-homogeneity of RF field.
کلیدواژه‌های انگلیسی مقاله MRI, RF homogeneity, Image uniformity, MR Coils

نویسندگان مقاله محمدعلی عقابیان | m a oghabian
associate professor, physics and biomedical engineering, tehran university of medical sciences, tehran, iran. research center for science amp;amp; technology in medicine, imam khomeini hospital, tehran. iran.

سازمان اصلی تایید شده: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)

sh مهدی پور | sh. mehdipour
m.sc. in research center for science amp;amp; technology in medicine, imam khomeini hospital, tehran. iran.


n riahicalam | n. riahicalam
associate professor, physics and biomedical engineering, tehran university of medical sciences, tehran, iran. research center for science amp;amp; technology in medicine, imam khomeini hospital, tehran. iran.

سازمان اصلی تایید شده: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)

b رفیعی | b. rafie
b. sc. in medical imaging center, imam khomeini hospital, tehran university of medical sciences, tehran, iran.

سازمان اصلی تایید شده: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)

h قناعتی | h. ghanaati
associate professor, radiology dept., medical imaging center, imam khomeini hospital, tehran university of medical sciences, tehran, iran.

سازمان اصلی تایید شده: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)


نشانی اینترنتی http://ijmp.mums.ac.ir/article_8124.html
فایل مقاله اشکال در دسترسی به فایل - ./files/site1/rds_journals/145/article-145-314858.pdf
کد مقاله (doi)
زبان مقاله منتشر شده en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات