Iranian Journal of Kidney Diseases
، جلد ۱۹، شماره ۰۴، صفحات ۲۳۲-۲۴۲
عنوان فارسی
چکیده فارسی مقاله
کلیدواژههای فارسی مقاله
عنوان انگلیسی
The Sensitivity of Bioimpedance to Assess Leg Fluid Volume Changes in Children on Hemodialysis: Experimental and Model-Based Investigation
چکیده انگلیسی مقاله
Introduction. Fluid management plays an essential role in pediatric hemodialysis; however, traditional assessment techniques often lack the precision required for this patient population. Bioimpedance analysis (BIA) offers a non-invasive alternative, but electrode placement significantly impacts measurement sensitivity. This study compared the efficacy of two calf-based electrode configurations — dorsal (ankle) versus plantar (sole) — in detecting fluid volume changes via bioimpedance in pediatric patients, particularly in terms of their sensitivity. Additionally, we evaluated a novel ring electrode arrangement for optimizing sensitivity distribution in lower limb tissues. Methods. This controlled before-and-after study was conducted at the dialysis unit of Ali Asghar Children’s hospital in Tehran, Iran. It evaluated thirteen children (aged 3-15 years) undergoing hemodialysis using bioimpedance analysis. The primary outcome measure was the change in impedance values before and after a single dialysis session. Measurements were performed with electrodes placed in two configurations: between the plantar surface of the foot to below the knee (down), and between the ankle dorsal surface to below the knee (up). Current distribution patterns were modeled using COMSOL simulations. Results. Mean ultrafiltration volume was 1123 ± 384 mL per session. Impedance changes were observed in both dorsal and plantar electrode configurations, with no significant differences or superiority in measuring these changes. Sensitivity analysis revealed variations of 35% with dorsal placement, 24% with plantar placement, and 20% with ring electrode configuration, demonstrating superior uniformity of the ring design across tissue regions. Conclusion. Optimal electrode placement varies among patients with no universal advantage for either position. Bioimpedance sensitivity depends on location of fluid accumulation, and enhanced sensitivity is required for monitoring changes below two liters. The ring electrodes provided the most uniform sensitivity distribution. These findings support use of personalized bioimpedance monitoring approaches to improve fluid management precision in pediatric hemodialysis.
کلیدواژههای انگلیسی مقاله
Diagnostic Techniques and Procedures, Blood Volume Determination, Electrodes, Dialysis, Bioelectrical impedance analysis, Pediatrics,Bioimpedance,Hemodialysis,Pediatric,Electrode Placement,Fluid Assessment
نویسندگان مقاله
| Shahrokh Noori
Department of Biomedical Engineering, Iranian Research Organization1 for Science and Technology (IROST), Tehran, Iran.
| Mohammad Firouzmand
Department of Biomedical Engineering, Iranian Research Organization1 for Science and Technology (IROST), Tehran, Iran.
| Vahid Reza Nafisi
Department of Biomedical Engineering, Iranian Research Organization1 for Science and Technology (IROST), Tehran, Iran.
| Nakysa Hooman
Professor of Pediatric Nephrology, Ali–Asghar Clinical Research Development Center, Department of Pediatrics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran.
نشانی اینترنتی
https://ijkd.org/index.php/ijkd/article/view/8556
فایل مقاله
فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
10.61186/
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
ORIGINAL | Dialysis
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات