Assessment of the effects of radiation type and energy on the calibration of TLD-100
چکیده انگلیسی مقاله
Introduction: In radiation therapy, knowing the dose rates to healthy organs and tumors is beneficial, and thermoluminescent dosimeter (TLD) allows for this possibility. This study was aimed at determining the dose-response differences of TLDs in various types of radiation, energy levels, and dose rate calibrated with other types of radiation beams and energy and dose levels. Materials and Methods: In this study, LiF:Mg,Ti (TLD-100) was used for dosimetry. Photon and electron irradiation was performed by Elekta Precise Linear Accelerator. First, TLDs were calibrated in three different groups of 6 MV photon, 6 MeV electron, and 60Co teletherapy photon beam with 50 cGy dose. Next, each group was irradiated with 6 MV photon, 6 MeV electron, and 60Co teletherapy photon beam separately at three different dose levels of 20, 60, and 100 cGy. Results: TLDs calibrated with electron were significantly different at all dose levels and with all types of radiation from TLDs calibrated with photon or 60Co teletherapy photon beam (P=0.000). P-value of the TLDs calibrated with 6 MV photon versus 60Co was less than 0.94. The maximum standard deviation belonged to 100 cGy irradiation, while the least pertained to 20 cGy irradiation. Conclusion: Calibration of TLDs depends on the type of radiation.
کلیدواژههای انگلیسی مقاله
Calibration, Dosimetry, Energy, Irradiation, TLD
نویسندگان مقاله
| Mohammad Taghi Bahreyni Toossi
Medical Physics Department, Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
| Hamid Gholamhosseinian
Medical Physics Research Centre, Faculty of medicine, Mashhad University of Medical Sciences, Mashhad, Iran
| Atefeh Vejdani Noghreiyan
Department of medical physics, Faculty of medicine, Mashhad University of Medical Sciences, Mashhad, Iran
نشانی اینترنتی
http://ijmp.mums.ac.ir/article_9933.html
فایل مقاله
اشکال در دسترسی به فایل - ./files/site1/rds_journals/145/article-145-658959.pdf
کد مقاله (doi)
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
Original Paper
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات