این سایت در حال حاضر پشتیبانی نمی شود و امکان دارد داده های نشریات بروز نباشند
صفحه اصلی
درباره پایگاه
فهرست سامانه ها
الزامات سامانه ها
فهرست سازمانی
تماس با ما
JCR 2016
جستجوی مقالات
سه شنبه 25 آذر 1404
Iranian Journal of Medical Physics
، جلد ۷، شماره ۱، صفحات ۴۳-۵۱
عنوان فارسی
چکیده فارسی مقاله
کلیدواژههای فارسی مقاله
عنوان انگلیسی
A Dosimetric Evaluation of Organs at Risk in Prostate Radiation Therapy using a MAGIC Gel Dosimeter
چکیده انگلیسی مقاله
Introduction: Multiple fields and presence of heterogeneities create complex dose distributions that need three dimensional dosimetry. In this work, we investigated MR-based MAGIC gel dosimetry as a three-dimensional dosimetry technique to measure the delivered dose to bladder and rectum in prostate radiation therapy. Materials and Methods: A heterogeneous slab phantom including bones was made. Paired cubes in the phantom representing bladder and prostate and a cylindrical container representing rectum were filled with MAGIC gel and placed in the anthropomorphic pelvic phantom. The phantom was irradiated with four beams as planned using a treatment planning system (TPS). Magnetic resonance transverse relaxation rate images were acquired and turned into dose distribution maps using a calibration curve. This calibration curve was obtained by linear fitting to R2 values of 4 test tubes against their given known doses. Image processing and data analysis were performed in MATLAB7 software. The gel dosimeter was validated using an ionization chamber. Dose maps and dose volume histograms (DVHs) were compared with dose distributions and DVHs of the TPS. Results: Mean “distance-to-agreement” and mean “dose difference” were 2.98 mm and 6.2%, respectively, in the comparison of profiles obtained from ionization chamber and gel dosimetry. Mean relative difference of DVHs between gel dosimetry and TPS data were 3.04%, 10.4% and 11.7%, for prostate, bladder and rectum, respectively. Discussion and Conclusions: Gel dosimetry is a good method for three dimensional dosimetry although it has a low precision in high dose gradient regions. This method can be used for evaluation of complicated dose distribution accuracy in 3D conformal radiotherapy, especially in presence of heterogeneities.
کلیدواژههای انگلیسی مقاله
نویسندگان مقاله
حمیدرضا صدوقی | hamid reza sadoughi
m.sc. student of medical physics, medical physics dept., tehran university of medical sciences, tehran, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)
محمود اله وردی | mahmood allahverdi
associate professor, medical physics dept., tehran university of medical sciences, and radiotherapy physics department of cancer institute, tehran, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)
حسن ندایی | hasan nedaei
assistant professor, radiotherapy physics dept., cancer institute, tehran, iran
رامین جابری | ramin jaberi
radiotherapy physicist, radiotherapy physics dept., cancer institute, tehran, iran
مهبد اصفهانی | mahbod esfahani
radiotherapy physicist, radiotherapy physics dept., cancer institute, tehran, iran
علیرضا شیرازی | alireza shirazi
associated professor, medical physics dept., tehran university of medical sciences, tehran, iran
سازمان اصلی تایید شده
: دانشگاه علوم پزشکی تهران (Tehran university of medical sciences)
نشانی اینترنتی
http://ijmp.mums.ac.ir/article_7270.html
فایل مقاله
فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده
en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
Original Paper
برگشت به:
صفحه اول پایگاه
|
نسخه مرتبط
|
نشریه مرتبط
|
فهرست نشریات