عنوان انگلیسی Standardization and optimization of Siemens Biograph TruePoint PET/CT acquisition and reconstruction parameters: Simultaneous qualitative and quantitative assessments
چکیده انگلیسی مقاله Introduction: Data acquisition and image reconstruction protocols affect image quality and quantification accuracy in PET imaging. We aimed to standardize and optimize image acquisition, and reconstruction parameter sets using a simultaneous quantitative and qualitative assessment framework for a lutetium oxyorthosilicate (LSO)-based PET/CT scanner.
Methods: The NEMA IEC Body Phantom acquisition was performed in list mode for 10 minutes with four spheres to background ratios (SBRs). Raw PET data were reconstructed using 60 different protocols. Image quality was evaluated for standardization using contrast, CNR, and noise. Recovery coefficient (RC) measurements were performed for different common VOI definitions.
Results: No significant differences were observed between RCs for various acquisition durations. The contrast to noise ratio (CNR) increased at all SBRs by expanding the acquisition duration from 60 to 600 seconds. PET scan time was reduced to 90 seconds per bed position while preserving image quality. Up to 50% improvement in CNR for the highest sub-iteration with a high level of smoothing was observed. PSF-based reconstruction produced a positive bias of RCmax in high SBRs (8 and 10) using higher sub-iterations (30 to 60) with Gaussian filters less than 6 mm FWHM. Moreover, a Sub-iteration of more than 32 with a 4-6 mm FWHM Gaussian filter provides optimized reconstruction sets.
Conclusion: Our study demonstrates it would be feasible for PET image acquisition and reconstruction settings to simultaneously allow optimal lesion detection with high image quality while providing accurate quantification.
کلیدواژه‌های انگلیسی مقاله PET/CT, Standardization, Optimization, Image quality

نویسندگان مقاله Habibeh Vosoughi |
Department of Medical Physics, Mashhad University of Medical Science, Mashhad, Iran

Farshad Emami |
Nuclear Medicine and Molecular Imaging Department, Imam Reza International University, Razavi Hospital, Mashhad, Iran

Mehdi Momennezhad |
Nuclear Medicine Research Center, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

Parham Geramifar |
Research Center for Nuclear Medicine, Tehran University of Medical Science, Tehran, Iran

Arman Rahmim |
Departments of Radiology and Physics, University of British Columbia, Vancouver, Canada


نشانی اینترنتی http://irjnm.tums.ac.ir/article_40166_ae11d1200fa2fa87272be3e51647316f.pdf
فایل مقاله فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات